News

 
  1. RTI Electronics AB is Looking for New Employees

    26 Jan 12 Would you like to be part of developing the medical technology products and markets of the future? RTI Electronics is growing and is now looking for new co-workers who want to be part of shaping the future in an exciting industry.

  2. RTI Electronics, Inc. has Moved

    02 Jan 12 RTI Electronics, Inc. has moved to larger location with lots of space to grow. Starting Tuesday, January 3, 2012 you can find RTI Electronics, Inc. at:

    33 Jacksonville Road
    Building 1
    Towaco, NJ 07082

    RTI Inc's phone numbers and contact information are the same for your convenience.

  3. Merry Christmas 2011 from RTI

    All the Best for 2012

    22 Dec 11 We wish to thank our customers and users for a fruitful 2011, wish you a really nice Christmas, and the best for 2012. Monday 26 December RTI Headquarters is closed.

  4. Ocean-Central-nov-2011webb-2.jpg

    Awaited RTI’s Ocean Central is Here

    07 Dec 11 Ocean Central makes it possible to tie all Ocean users together via network and internet. With Ocean Central you can save all measured data in one place and easily distribute templates to all users.

Show all news
Chamber Adapter

DCT10 Pencil Ionization Chamber

DCT10 Pencil Ionization Chamber

Ionization Chamber for CTDI Measurements

The DCT10 RS is rugged, cylindrical pencil shaped air ionization chamber for CTDI measurements. The CT probe is intended for measuring and monitoring the exposure output level of CT scanners in a phantom or in air. CT dose index, CTDI, can be measured in accordance with IEC 61223-2-6.

 


DCT10 and Piranha

The DCT10 RS can be connected via Chamber Adapter to the Piranha to measure CT dose.

Ranges with Piranha (Chamber Adapter required)
Dose Rate 80 µGycm/s to 8 Gycm/s
Inaccuracy ±6 % or ±0.080 mGycm/s

 

DCT10 and the Barracuda

The DCT10 RS can be connected directly to the Barracuda (via an electrometer module supporting bias) to measure CT dose. Barracuda has built-in functions to simplify measurements of CTDI (CT dose index).

Measuring ranges with Barracuda (with EMM-Biasw)
Dose 40 nGym-350 kGym
4 µGycm-35 MGycm

Dose rate 1.6 µGym/s-70 kGym/s
160 µGycm/s-7kGycm/s
Inaccuracy ±5 % or 16 µGycm/s
 
DCT10 and the Solidose Family

The DCT10 RS can be connected directly to the Solidose 400 to measure CT dose.

Measuring ranges Solidose
Dose 20 µGycm - 400 kGycm ±5 %
2.3 mRcm - 46 MRcm
Dose rate 200 µGycm/s - 600 Gycm/s ±5 %
1.4 Rcm/min - 4 MRcm/min

General Specifications DCT10 Pencil Ionization Chamber
Connector type
LEMO triaxial
Cable
2 m, low noise triaxial
Active volume
4.9 cm3
Active length
100 mm
Inner diameter
8.0 mm
Typical leakage
±4 fA
Radiation quality 80 - 150 kV
Sensitivity
40 nC/Gycm or 400 nC/Gy
  0.35 nC/Rcm or 3.5 nC/R
 
 