News

 
  1. RTI Electronics AB is Looking for New Employees

    26 Jan 12 Would you like to be part of developing the medical technology products and markets of the future? RTI Electronics is growing and is now looking for new co-workers who want to be part of shaping the future in an exciting industry.

  2. RTI Electronics, Inc. has Moved

    02 Jan 12 RTI Electronics, Inc. has moved to larger location with lots of space to grow. Starting Tuesday, January 3, 2012 you can find RTI Electronics, Inc. at:

    33 Jacksonville Road
    Building 1
    Towaco, NJ 07082

    RTI Inc's phone numbers and contact information are the same for your convenience.

  3. Merry Christmas 2011 from RTI

    All the Best for 2012

    22 Dec 11 We wish to thank our customers and users for a fruitful 2011, wish you a really nice Christmas, and the best for 2012. Monday 26 December RTI Headquarters is closed.

  4. Ocean-Central-nov-2011webb-2.jpg

    Awaited RTI’s Ocean Central is Here

    07 Dec 11 Ocean Central makes it possible to tie all Ocean users together via network and internet. With Ocean Central you can save all measured data in one place and easily distribute templates to all users.

Show all news

Piranha Panoramic Holder (OPG)

Piranha Panoramic Holder (OPG)

Field Position Analyzer that Saves Time and Money

The Piranha is ideal for measurement on panoramic (OPG) systems. The small detector area and the "Piranha position check", a feature that makes it possible to verify the detector position, ensures more accurate and more reproducible results than with other devices, since the Piranha works wireless, you do not have to worry about any cable during the rotation of the OPG. The Piranha OPG holder makes it possible to attach the Piranha on the panoramic system and quickly perform the measurement. Of course, the Piranha is perfectly suited for measurements on standard intra oral systems as well.


The Piranha Panoramic Holder can be fitted with magnets or adhesive tape. The holder comes with several bendable plates that can be used to fit "around corners" depending on the geometry of the panoramic (OPG ) system. Two screws makes it possible to adjust the horizontal position +/-2 mm.

This means that the holder doesn't need to be positioned "perfectly" with the magnets or tape but can be adjusted afterwards. Note that on some panoramic systems, magnets are not allowed or not possible to use, and the plates makes it easy to use adhesive tape instead.

To perform a measurement:
  1. Attach the holder and MPD.
  2. Select the "MPD Position Check" and make an exposure.
  3. Select "measure all".
  4. Make an exposure and get kVp, exposure time, dose, dose rate, HVL total filtration, and waveform. All in one exposure!
 
 